Introduction
-AR-Y200ventilator is a new-type of pneumatic-driven and electronic-controlledventilator which is designed to meet the special needs of clinical respiratory managementand respiratory therapy for infants and children with respiratory disorders.With the feature of innovative functions, convenient operation, less gasconsumption, compact but reasonable structure, and outstanding safety andreliability, it is an indispensable instrument in emergency and ICU rooms, andan essential tool to rescue the life of infants and children.
-AR-Y200 ventilatorcan work with some innovative ventilation modes, e.g., constant flow
mode,pressure limit mode and time cycle ventilation mode, and also supports PEEP andIMV functions.
TechnicalSpecifications
PARAMETER | SPECIFICATION | |
Working Mode (Ventilation Mode) | Controlled Mechanical Ventilation (CMV), Intermittent Mandatory Ventilation (IMV), Continuous Positive Airway Pressure (CPAP), Manual Ventilation (Manual). | |
Breath Rate | 1~120BPM | |
Inspiratory to Expiratory Ratio | 1:1~1:3 | |
Special Order I:E Range | 10:1~1:9.9 | |
Inspiratory Flow | 3 L/min~20 L/min | |
Tidal Volume | Minimum 10 mL~300 mL | |
Minute Volume | Max. Ventilation volume greater than 5 L/min | |
Oxygen Concentration of Output Gas | 21%~100% | |
Ventilator System Compliance | Not greater than 3×10-2 mL/Pa | |
Positive End-Expiratory Pressure (PEEP) | 0kPa~2.0kPa | |
Inhalation Safety Valve | Safety valve shall be opened when the airway pressureis within -300 Pa ~ -800 Pa | |
Inspiratory Pressure | 1kPa~9kPa | |
Inspiratory Time | 0.2s~3.0s | |
Synchronous Trigger | Pressure Trigger | -1.0~0kPa (Special Order) |
Flow Trigger | 0~10L (Special Order) | |
Maximum Safe Pressure | Not greater than 10kPa | |
Continuous Positive Airway Pressure (CPAP) | 0kPa~2.0kPa | |
Monitoring Tidal Volume | Real-time monitoring within the adjustment range oftidal volume: <100 mL: ±10 mL; ≥100 mL: ±15 %; | |
Monitoring Oxygen Concentration | 21%~100% | |
Power Failure Alarm | Audible alarm sounds immediately in case of power outageand lasts at least for 120s | |
Gas Source Failure Alarm | Audible and visual alarm sounds within 6s after gas sourceinterruption occurs and lasts at least for 120s | |
Airway Pressure Lower Limit Alarm Range | PEEP +0.2kPa | |
Airway Pressure Upper Limit Alarm Range | Inspiratory pressure +0.4 kPa |